HPLC purity vs identity: what each measures, and what neither proves
An HPLC purity figure is the target peptide's share of the peptide signal a detector sees in a chromatogram. An identity test asks a different question: is this the peptide named on the label? A sample can score well on one and fail the other. Neither measures how much peptide the vial holds, or anything about microbial contamination.
Both rows sit near the top of every certificate we publish, and both are easy to read as a single verdict on quality. They are not. This note takes each in turn, using the certificates for two of our own lots: BPC-157, certificate reference 030284, tested by ILS Laboratories, and SS-31, certificate reference 060722, tested by BTLabs.
What does an HPLC purity figure measure?
It measures the main peak's share of the total peak area the detector records at one wavelength: a relative figure, not an amount.
High-performance liquid chromatography pushes a dissolved sample through a packed column. Different molecules travel through it at different speeds and leave it at different times, so the detector at the end records a series of peaks over time. For peptides the workhorse is reversed-phase HPLC, which separates molecules mainly by how hydrophobic they are; ion-exchange and size-exclusion modes are also used[1].
The detector most purity methods use reads ultraviolet absorbance. A purity percentage is then a ratio of areas: the area of the main peak divided by the total area of the peaks counted. The ILS Laboratories certificate for 030284 says exactly that in its notes: purity is "determined by RP-HPLC area normalization", and the value "represents the percentage of the target peptide relative to all peptide-related peaks". On that certificate the figure is 99.72%, by RP-HPLC at 214 nm. The BTLabs certificate for 060722 gives 99.6%, by HPLC at 280 nm.
So a purity figure is relative. It answers: of everything this detector saw at this wavelength, how much was the main peak?
Why does the detection wavelength matter?
Because different parts of a peptide absorb different amounts of light, and the amount changes with wavelength.
Kuipers and Gruppen measured how strongly each amino acid and the peptide bond absorb at 214 nm, a wavelength where the peptide backbone itself absorbs[2]. The peptide bond's molar extinction coefficient was 923 M-1 cm-1. Tryptophan absorbed about 30 times more than a peptide bond, and phenylalanine, tyrosine and histidine about 6 times more; proline inside a chain absorbed about 3 times more[2]. Two peptides of the same length can therefore give different peak areas for the same amount of material, and an impurity's share of the total area depends on its composition as well as on how much of it there is.
That is why purity is reported with its wavelength, and why ICH Q6B warns that purity figures "are highly method-dependent"[4]. Our two laboratories read at different wavelengths: ILS Laboratories at 214 nm and BTLabs at 280 nm on the two lots above. Each figure is correct for its own method. They are not two readings of the same scale.
What can a purity figure miss?
A purity percentage is only as complete as the peaks it counts.
- Impurities that are not peptides. The ILS Laboratories certificate states that "non-peptide process-related impurities, if detected, are excluded from the calculation". Its purity figure is a purity among peptide-related peaks by design.
- Counter-ions and water. Synthetic peptides commonly carry counter-ions such as trifluoroacetate from synthesis or purification[3]. These add weight to the powder without adding a peptide peak, so they lower the peptide's share of the mass without lowering its chromatographic purity. That is a question for content, not purity.
- Impurities that run with the main peak. Solid-phase synthesis, the route by which most peptides are made, produces related impurities: sequences missing an amino acid or carrying an extra one, diastereomers from racemization, side-chain adducts, oxidation products, dimers and larger aggregates[3]. A related impurity that elutes at the same time as the target adds to the main peak's area rather than appearing as a separate peak.
- Anything the detector does not see. A compound that absorbs little at the chosen wavelength contributes little area, whatever its mass.
D'Hondt and colleagues point out why this matters even outside medicine: related impurities can influence early functional studies and lead to erroneous conclusions[3]. For an in-vitro researcher the purity figure is a necessary number, not a sufficient one.
What does an identity test establish?
Identity is a yes-or-no question about the molecule itself. ICH Q6B asks that identity tests be "highly specific for the drug substance" and based on "unique aspects of its molecular structure and/or other specific properties"[4]. ICH Q2(R2) adds that an identification test should also be challenged with materials structurally similar or closely related to the analyte, because a test that cannot tell near neighbours apart has not shown much[5].
Our two laboratories use different identity methods on our lots:
- ILS Laboratories: HPLC retention time vs. reference standard. For 030284 the row reads "BPC-157, Confirmed".
- BTLabs: FTIR (USP <197A>). For 060722 the row reads "SS-31, Confirmed".
How the two methods work, and what each can miss, is the subject of retention time vs FTIR.
Why can a sample pass one test and fail the other?
Because they measure independent properties.
- High purity, wrong identity. A clean preparation of the wrong peptide gives a single dominant peak and a high purity percentage. Purity has no way to know which molecule the peak is; only an identity test does.
- Right identity, lower purity. A preparation of the right peptide that carries truncated sequences or oxidation products will show the expected identity and a lower purity percentage.
- Right identity, high purity, less peptide than the label. Neither test weighs the peptide. A sample can be the right molecule, chromatographically clean, and still contain less peptide per vial than its label because of counter-ions, water or fill.
This independence is why ICH Q2(R2) recommends that "where one analytical procedure does not provide sufficient discrimination, a combination of two or more procedures" be used[5], and why a certificate carries both rows.
What does each row answer, and what does it leave open?
Purity answers what share of the signal is the main peak, identity whether the material is the named peptide, content how much peptide was measured, and endotoxin and sterility the microbial questions; none answers the others.
| Row on the certificate | The question it answers | What it leaves open |
|---|---|---|
| Purity (RP-HPLC at 214 nm at ILS Laboratories) | What share of the peptide signal at this wavelength is the main peak? | Which molecule the main peak is; how much peptide is present; non-peptide impurities |
| Identity (HPLC retention time vs. reference standard at ILS Laboratories; FTIR (USP <197A>) at BTLabs) | Is the material the named peptide, by this method? | How clean it is; how much of it there is |
| Content (Report Only at ILS Laboratories; 90 – 110 % of 10 mg at BTLabs) | How much peptide was measured in the sample? | Purity and identity, which are separate rows |
| Endotoxin and sterility | Was bacterial endotoxin measured, and did the sample show microbial growth? | Everything chemical about the peptide |
What does neither test prove?
Neither proves how much peptide is in the vial, its microbial quality, anything about the individual vial in your hands, or its suitability for a particular assay.
- How much peptide is in the vial. That is the content row, discussed in Report Only and net peptide content.
- Microbial quality. The BTLabs certificate says so about itself: "It does not provide information on particulate matter, microbial contamination or presence of endotoxins." Endotoxin and sterility are separate tests, discussed in endotoxin and sterility.
- The individual vial in your hands. A certificate describes the sample the laboratory received; our testing standard explains what that does and does not cover.
- Suitability for a particular assay. Whether a figure is good enough is a judgement for the researcher's own method, not a property of the certificate.
Our own release standard sets a floor on the purity row, ≥99% HPLC purity, and requires the identity row the laboratory reports. Each released lot's certificate is on its product page and on the certificates page, and the laboratories page lists the methods each laboratory uses on our lots.
Every figure on this page that describes an Elyria Bio lot is generated from the same certificate records that feed the product pages and the certificates page, and is read from the laboratory’s own certificate. What a lot must meet before release is set out in our testing standard; who tests our lots, and how, is on the laboratories page. Any other statement of fact cites its source in the list below.
All Elyria Bio products are Research Use Only (RUO) in-vitro laboratory reference materials. They are not for human or veterinary consumption and are not drugs, foods, cosmetics, or dietary supplements. Nothing on this site is medical advice or a dosing recommendation. Purchasing or using these materials in any way that violates this policy or applicable law is strictly prohibited.
Sources
- HPLC analysis and purification of peptides (Methods Mol Biol, 2007). Mant CT, Chen Y, Yan Z, et al.; PubMed 18604941. Link and title checked 2026-10-23.
- Prediction of molar extinction coefficients of proteins and peptides using UV absorption of the constituent amino acids at 214 nm (J Agric Food Chem, 2007). Kuipers BJ, Gruppen H; PubMed 17539659. Link and title checked 2026-10-23.
- Related impurities in peptide medicines (J Pharm Biomed Anal, 2014). D'Hondt M, Bracke N, Taevernier L, et al.; PubMed 25044089. Link and title checked 2026-10-23.
- ICH Q6B: Specifications: Test Procedures and Acceptance Criteria for Biotechnological/Biological Products. International Council for Harmonisation (ICH). Link and title checked 2026-10-23.
- ICH Q2(R2): Validation of Analytical Procedures. International Council for Harmonisation (ICH). Link and title checked 2026-10-23.