HPLC retention time vs FTIR for identity: what our two laboratories do differently

Both are identity tests, but they ask different questions. HPLC retention time asks whether the sample's main peak leaves the column when a reference standard's does, under the same conditions. FTIR asks whether the sample's infrared spectrum shows the absorption bands expected of the named peptide. On our lots, ILS Laboratories uses the first and BTLabs the second.

Both laboratories report the identity of every lot we release, and both report it as "Confirmed". The word is the same; the evidence behind it is not. This note explains what each method compares, what each can miss, and how to read the identity row on our certificates. It uses the ILS Laboratories certificate for BPC-157 (certificate reference 030284) and the BTLabs certificate for SS-31 (certificate reference 060722).

What makes a good identity test?

A good identity test is highly specific[2], and shows it by telling the named molecule apart from materials structurally similar to it[1].

Regulatory guidance sets the bar clearly. ICH Q6B asks that identity tests be "highly specific" and based on "unique aspects of its molecular structure and/or other specific properties"[2]. ICH Q2(R2) describes how that specificity is shown: a positive finding comparable to a reference material, a negative finding from samples that do not contain the analyte, and a challenge with materials that are structurally similar or closely related to it[1]. The last part is the demanding one. Telling a peptide apart from water is easy; telling it apart from a near neighbour is the test.

Every identity method has some neighbours it cannot separate. That is why the same guidance recommends that "where one analytical procedure does not provide sufficient discrimination, a combination of two or more procedures" be used[1].

What does HPLC retention time compare?

It compares the time the sample's peak takes to leave the column with the time a reference standard's peak takes under the same conditions.

In reversed-phase HPLC a peptide is carried through a column by a changing solvent mixture and retained according to how strongly it interacts with the column's surface, which for peptides is mainly a matter of hydrophobicity[3]. The time from the moment the sample enters the column to the top of its peak is its retention time.

A retention time on its own identifies nothing, because it depends on the column, the solvents, the gradient and even the column temperature: Villacrés and colleagues measured peptide retention shifting steadily as column temperature changed, and traced the shift to each peptide's residues, hydrophobicity, length and helicity[4]. What makes retention time an identity test is a comparison. The laboratory runs a reference standard of the named peptide under the same conditions and asks whether the sample's main peak appears at the same time.

That is what the ILS Laboratories certificate for 030284 states in its notes: "The sample was confirmed to be BPC-157 by HPLC. Identification by chromatographic retention time comparison with a reference standard." The identity row reads "BPC-157, Confirmed", by HPLC retention time vs. reference standard.

What it can miss. Two different molecules can elute at nearly the same time under one set of conditions. A sequence variant that differs by a single, similar residue, or a molecule with similar overall hydrophobicity, may sit close to the target. The comparison is strongest when the reference standard is well characterised and the separation is good enough to resolve near neighbours; it is weakest when it is not challenged with them.

What does an FTIR spectrum compare?

It compares the infrared bands a sample absorbs with the characteristic bands expected of the peptide.

Fourier transform infrared spectroscopy shines infrared light through or onto a sample and records which wavelengths the sample absorbs. Bonds vibrate at characteristic frequencies, so the spectrum is a fingerprint of the chemical groups present and of how the molecule is folded. It is a long-established, non-destructive technique for polypeptides and proteins, and the bands it records are correlated with secondary-structure components[5].

On our lots, BTLabs reports identity by FTIR (USP <197A>). Its certificate for 060722 sets the specification "Sample spectrum should confirm the content of peptide via characteristic bands", and reports: "FTIR sample spectrum confirms the presence of SS-31 with addition of excipient(s)/fillers." The identity row reads "SS-31, Confirmed".

What it can miss. A spectrum is dominated by the groups a molecule has most of, and every peptide shares the same backbone. Peptides of similar composition can give similar spectra. The vial's other contents add bands of their own, which is why the certificate names "excipient(s)/fillers" and why the same certificate weighs the whole contents (100.5 mg) and gives a peptide-to-excipient ratio of 10.4 : 90.1 mg (1:8.6). An FTIR spectrum confirms the presence of the expected bands; it is a weaker tool for telling near neighbours apart.

How do the two laboratories differ on our lots?

ILS Laboratories confirms identity by HPLC retention time vs. reference standard; BTLabs by FTIR (USP <197A>).

The table below is generated from every lot we have released, grouped by the laboratory that tested it. Each method string is printed exactly as the laboratory's certificate prints it.

LaboratoryReleased lotsPurity method, as printedLaboratory purity specificationIdentity method, as printed
ILS Laboratories12RP-HPLC at 214 nm≥95.0%HPLC retention time vs. reference standard
BTLabs5HPLC at 220 nm; HPLC at 280 nm98%FTIR (USP <197A>)

Three differences follow from it:

  • The identity method. ILS Laboratories identifies by HPLC retention time vs. reference standard; BTLabs by FTIR (USP <197A>). They are different measurement principles.
  • The purity wavelength. ILS Laboratories reads purity at 214 nm on our lots and BTLabs at 280 nm, which matters for purity and is discussed in HPLC purity vs identity.
  • The chromatogram detail. The BTLabs certificate prints each HPLC peak's retention time and area. For 060722 its first peak is at 3.913 min with an area of 1318.4 mAU*s. That retention time belongs to the column and conditions BTLabs used; it is not comparable with a retention time from another laboratory's method.

Which method is better?

Neither, in general. They are orthogonal: they rest on different physical principles and fail in different ways. ICH Q2(R2) treats comparison with "a second, well characterised analytical procedure that ideally applies a different measurement principle" as one way to verify specificity[1]. A retention-time match says the sample behaves like the reference standard on one column; an FTIR match says it carries the expected chemical groups. Each covers some of the other's blind spots.

What neither does on its own is sequence the peptide or measure its exact mass. When a certificate's identity row is the only identity evidence, read it as what it is: one method's finding, against that method's specification.

What should you ask when identity matters most to your work?

Some in-vitro work depends on telling a peptide apart from a close analogue: a study that compares two related sequences, for example, or one where a truncated variant would confound the read-out. There the useful questions follow directly from the guidance above.

  • Which method was used, and against what? A retention-time comparison is only as good as its reference standard and its separation; an FTIR comparison is only as good as the bands it names[1].
  • Was the method challenged with near neighbours? ICH Q2(R2) asks that an identification test also be applied to materials that are structurally similar or closely related to the analyte[1]. A laboratory can tell you whether its method was.
  • Would a second principle add anything? Where one procedure does not discriminate enough, the same guidance recommends a combination of two or more[1]. Retention time and FTIR rest on different principles, so each is a natural complement to the other.

None of these questions is answered on the certificate itself. They are questions for the laboratory, and the laboratories page lists how to reach each one's record.

How should you read the identity row on our certificates?

Read the method the row names first: the row's result means what that method can establish, and no more.

  1. Find the method the row names. On our lots it is either HPLC retention time vs. reference standard or FTIR (USP <197A>); the laboratories page lists which laboratory tested which lot.
  2. Read the specification the laboratory set, where it prints one. For FTIR on the 060722 certificate it is the characteristic-bands sentence quoted above.
  3. Read the finding as the laboratory words it. "Confirmed" means the method's criterion was met, not that every possible alternative was excluded.
  4. Remember what the row does not cover: purity, content and microbial quality are separate rows, each explained in how to read a certificate of analysis.

Every released lot's identity row is published as the laboratory printed it, on its product page and on the certificates page.

About the data

Every figure on this page that describes an Elyria Bio lot is generated from the same certificate records that feed the product pages and the certificates page, and is read from the laboratory’s own certificate. What a lot must meet before release is set out in our testing standard; who tests our lots, and how, is on the laboratories page. Any other statement of fact cites its source in the list below.

Research use only

All Elyria Bio products are Research Use Only (RUO) in-vitro laboratory reference materials. They are not for human or veterinary consumption and are not drugs, foods, cosmetics, or dietary supplements. Nothing on this site is medical advice or a dosing recommendation. Purchasing or using these materials in any way that violates this policy or applicable law is strictly prohibited.

Sources

  1. ICH Q2(R2): Validation of Analytical Procedures. International Council for Harmonisation (ICH). Link and title checked 2026-10-23.
  2. ICH Q6B: Specifications: Test Procedures and Acceptance Criteria for Biotechnological/Biological Products. International Council for Harmonisation (ICH). Link and title checked 2026-10-23.
  3. HPLC analysis and purification of peptides (Methods Mol Biol, 2007). Mant CT, Chen Y, Yan Z, et al.; PubMed 18604941. Link and title checked 2026-10-23.
  4. Toward an Ultimate Solution for Peptide Retention Time Prediction: The Effect of Column Temperature on Separation Selectivity (J Proteome Res, 2024). Villacrés C, Mizero B, Spicer V, et al.; PubMed 38530092. Link and title checked 2026-10-23.
  5. Fourier transform infrared spectroscopic analysis of protein secondary structures (Acta Biochim Biophys Sin, 2007). Kong J, Yu S; PubMed 17687489. Link and title checked 2026-10-23.