How to read a peptide certificate of analysis, row by row
A certificate of analysis is a laboratory's report on one sample drawn from one batch. Read it in order: who issued it and for which lot; purity, the target's share of the detected peptide signal; identity, whether it is the named peptide; content, how much peptide the vial holds; then any microbial tests. Each row answers a different question.
This note reads two real certificates for our own lots, one from each laboratory that tests them, line by line: the ILS Laboratories certificate for BPC-157, certificate reference 030284, and the BTLabs certificate for SS-31, certificate reference 060722. Both are public. The BPC-157 certificate is linked as a PDF from the product page, and the SS-31 lot has a page of its own with each of its reports. Open one beside this page and follow along.
What is a certificate of analysis, and what is it for?
A certificate of analysis records what a laboratory measured on a sample and against which criteria. Regulators use a precise word for the criteria: ICH Q6B defines a specification as "a list of tests, references to analytical procedures, and appropriate acceptance criteria which are numerical limits, ranges, or other criteria for the tests described"[1]. A certificate is therefore three things side by side on every row: the test, the method the laboratory used, and either a limit the figure is held to or a note that no limit applies.
The tests on a peptide certificate fall into a few families that the pharmacopoeias recognise for peptides: identification, purity and assay of content[3]. Vergote and colleagues, comparing the published monographs, found that even the pharmacopoeias do not agree on how each family is tested or where its limits sit[3]. That is the first reason to read the method column as carefully as the figure: two laboratories can both report "purity" and mean measurably different things.
Who issued the certificate, and for which lot?
The header answers both: it prints the issuing laboratory's name and address, the report number, and the lot number this site calls the certificate reference.
Start at the top. The header tells you whose document this is and what it describes, and every later row depends on it.
- The laboratory. The BPC-157 certificate is issued by ILS Laboratories (San Diego CA); the SS-31 certificate by BTLabs (West Palm Beach FL). Each prints its own name and address on its own letterhead.
- The report number. The laboratory's own identifier for the document: COA-2026-955273 for BPC-157 and COA261149-004 for SS-31. This is the number to quote to the laboratory.
- The lot. Both certificates print a "Lot Number" line. That number is what this site calls the certificate reference: 030284 and 060722. The BTLabs certificate also prints a sample ID, the laboratory's internal tag for the sample it received.
- The labelled content. The amount the vial is labelled to hold, as the certificate prints it: 10 mg for BPC-157 and 10 mg for SS-31.
- The date. The analysis date: 2026-08-07 for 030284 and 2026-08-27 for 060722.
The ILS Laboratories certificate also prints "Scan to verify authenticity at ils-lab.com", an access code (4FT5KXQ2) and a verification link. Those are what let you retrieve the laboratory's own copy, which the last note in this series walks through[4].
What does the purity row say?
On the ILS Laboratories certificate the row reads "Peptide Purity (HPLC)" with a specification of ≥95.0% and a result of 99.72%. The method, as the certificate prints it, is RP-HPLC at 214 nm. The certificate's own notes explain what the percentage is: "Value represents the percentage of the target peptide relative to all peptide-related peaks." It is a share of the chromatogram, not a weight.
On the BTLabs certificate the HPLC purity row gives a specification of 98% and a result of 99.6%, measured by HPLC at 280 nm.
Three things are worth noticing. The specification on each row is the laboratory's, not ours. Our own release floor, ≥99% HPLC purity, is stricter than either and is set out in our testing standard. And the two laboratories read their chromatograms at different wavelengths, which is one reason a purity figure is "highly method-dependent", in ICH Q6B's words[1]. What a purity figure does and does not cover is the subject of HPLC purity vs identity.
What does the identity row say?
It says whether the sample is the named molecule, by the method the row names: HPLC retention time vs. reference standard at ILS Laboratories, FTIR (USP <197A>) at BTLabs.
Purity tells you how clean the sample is. Identity tells you whether it is the right molecule, and ICH Q6B asks that an identity test be "highly specific" and based on "unique aspects of its molecular structure"[1].
- ILS Laboratories: identity by HPLC retention time vs. reference standard, reported as "BPC-157, Confirmed".
- BTLabs: identity by FTIR (USP <197A>), reported as "SS-31, Confirmed".
The two methods are different kinds of evidence. One compares how long the sample takes to pass through a chromatography column with a reference standard; the other compares the sample's infrared spectrum with the bands expected of the peptide. ICH Q2(R2) describes the evidence an identification test should produce: a positive finding comparable to a reference material, and a negative one from samples that do not contain the analyte[2]. How the two methods differ, and what each can miss, is in retention time vs FTIR.
What does the content row say?
It reports how much peptide the laboratory measured in the sample: a mass, not a purity.
This is the row readers most often misread. It is not purity; it is mass.
On the ILS Laboratories certificate the row is "Net Peptide Content", and its specification column reads Report Only. The result is 10.7 mg against a 10 mg label. Our record of the row says: Reported as Net Peptide Content against a Report Only specification; the certificate sets no acceptance range.
On the BTLabs certificate the equivalent row is the HPLC potency assay, and it does carry a range: 90 – 110 % of 10 mg. The result is 10.4 mg (104.5 %). The same certificate weighs the whole contents of the vial (100.5 mg) and gives a peptide-to-excipient ratio of 10.4 : 90.1 mg (1:8.6), with the note "USP recommended ratios of (1:0) to (1:10)".
Why a vial can show more peptide than its label, and what "Report Only" does and does not mean, is the subject of Report Only and net peptide content.
Where are the endotoxin and sterility tests?
ILS Laboratories prints them on the same certificate as purity; BTLabs issues them as separate reports with their own numbers.
The two laboratories place them differently, and the difference matters when you look for them.
- ILS Laboratories reports both on the same certificate as purity. Endotoxin by USP <85>: 0.101 EU/mL, status "Reported". Sterility by PCR: No Growth, PASS.
- BTLabs reports them on separate reports with their own numbers: endotoxin on COA261149-018 (9.9 EU/vial (limit 350 EU/vial, 2 mL reconstituted), PASS) and sterility on COA261149-011 (No growth, PASS).
The main BTLabs certificate says so in its own words: "It does not provide information on particulate matter, microbial contamination or presence of endotoxins." A purity certificate on its own therefore tells you nothing about microbial quality; look for the separate reports. What each test is, and why some panels carry it and others do not, is in endotoxin and sterility.
The two certificates side by side
| Row | ILS Laboratories, BPC-157 (030284) | BTLabs, SS-31 (060722) |
|---|---|---|
| Report number | COA-2026-955273 | COA261149-004 |
| Analysis date | 2026-08-07 | 2026-08-27 |
| Purity, as printed | 99.72% (specification ≥95.0%) | 99.6% (specification 98%) |
| Purity method | RP-HPLC at 214 nm | HPLC at 280 nm |
| Identity | BPC-157, Confirmed, by HPLC retention time vs. reference standard | SS-31, Confirmed, by FTIR (USP <197A>) |
| Content | 10.7 mg (Report Only) | 10.4 mg (104.5 %) (90 – 110 % of 10 mg) |
| Endotoxin | 0.101 EU/mL, Reported | 9.9 EU/vial (limit 350 EU/vial, 2 mL reconstituted), PASS |
| Sterility | No Growth, PASS | No growth, PASS |
What does a certificate not tell you?
It does not describe your individual vial, a test that was not ordered, or whether a "Report Only" figure passes.
A careful reading ends with the limits of the document.
- It describes the sample the laboratory received. The certificate is evidence about the batch that sample was drawn from. Our vials are not individually lot-numbered, so a certificate cannot be matched to one individual vial, and nothing we publish says it can (see our testing standard).
- It reports only the tests that were ordered. A row that is not on the certificate was not tested on that document. The BTLabs exclusion sentence above is the clearest example.
- A "Report Only" figure is a measurement, not a pass. Read the specification column before reading the result as good or bad.
How do you check that a certificate is genuine?
Retrieve the laboratory's own copy and compare it with yours, row by row.
Reading a certificate assumes it is the laboratory's document. The way to confirm that is to retrieve the laboratory's own copy and compare the two, row by row. ILS Laboratories publishes each of its certificates on a portal reached by the access code printed on the certificate[4]; BTLabs links each report to a verification copy from our lot pages. How to verify a certificate with the laboratory walks through both, step by step, and the laboratories page lists every lot each laboratory has tested.
Every figure on this page that describes an Elyria Bio lot is generated from the same certificate records that feed the product pages and the certificates page, and is read from the laboratory’s own certificate. What a lot must meet before release is set out in our testing standard; who tests our lots, and how, is on the laboratories page. Any other statement of fact cites its source in the list below.
All Elyria Bio products are Research Use Only (RUO) in-vitro laboratory reference materials. They are not for human or veterinary consumption and are not drugs, foods, cosmetics, or dietary supplements. Nothing on this site is medical advice or a dosing recommendation. Purchasing or using these materials in any way that violates this policy or applicable law is strictly prohibited.
Sources
- ICH Q6B: Specifications: Test Procedures and Acceptance Criteria for Biotechnological/Biological Products. International Council for Harmonisation (ICH). Link and title checked 2026-10-23.